Accuracy of inpatient oxygen administration.
AUTOR(ES)
Jeffrey, A A
RESUMO
The accuracy of supplemental oxygen delivery was assessed in two medical units. Only 90 of 206 patients (44%) were receiving oxygen as prescribed. The ward oxygen rotameters were inaccurate at normal clinical flow rates, the range of flows being delivered as a percentage of the indicated flow varying from 15% at 8 l/min to 40% at 1 and 2 l/min.