An in-house real-time polymerase chain reaction: standardisation and comparison with the Cobas Amplicor HBV monitor and Cobas AmpliPrep/Cobas TaqMan HBV tests for the quantification of hepatitis B virus DNA
AUTOR(ES)
Santos, Ana Paula de Torres, Levi, José Eduardo, Lemos, Marcilio Figueiredo, Calux, Samira Julien, Oba, Isabel Takano, Moreira, Regina Célia
FONTE
Mem. Inst. Oswaldo Cruz
DATA DE PUBLICAÇÃO
2016-02
RESUMO
This study aimed to standardise an in-house real-time polymerase chain reaction (rtPCR) to allow quantification of hepatitis B virus (HBV) DNA in serum or plasma samples, and to compare this method with two commercial assays, the Cobas Amplicor HBV monitor and the Cobas AmpliPrep/Cobas TaqMan HBV test. Samples from 397 patients from the state of São Paulo were analysed by all three methods. Fifty-two samples were from patients who were human immunodeficiency virus and hepatitis C virus positive, but HBV negative. Genotypes were characterised, and the viral load was measure in each sample. The in-house rtPCR showed an excellent success rate compared with commercial tests; inter-assay and intra-assay coefficients correlated with commercial tests (r = 0.96 and r = 0.913, p < 0.001) and the in-house test showed no genotype-dependent differences in detection and quantification rates. The in-house assay tested in this study could be used for screening and quantifying HBV DNA in order to monitor patients during therapy.
Documentos Relacionados
- Quantification of Human Immunodeficiency Virus Type 1 Proviral DNA by the TaqMan Real-Time PCR Assay†
- Quantification of Human Immunodeficiency Virus Type 1 Proviral Load by a TaqMan Real-Time PCR Assay
- Real-Time TaqMan PCR for Quantifying Oral Bacteria during Biofilm Formation
- Detection and Quantification of Methyl tert-Butyl Ether-Degrading Strain PM1 by Real-Time TaqMan PCR
- In-house quantitative real-time PCR for the diagnosis of hepatitis B virus and hepatitis C virus infections