Assessment of Laboratory Performance of Nucleic Acid Amplification Tests for Detection of Mycobacterium tuberculosis
AUTOR(ES)
Ridderhof, John C.
FONTE
American Society for Microbiology
RESUMO
During implementation of the Centers for Disease Control and Prevention's Mycobacterium tuberculosis nucleic acid amplification (NAA) evaluation program, 27.1% of participants used the same biological safety cabinet for NAA and specimen processing; 28.8% reported not using unidirectional workflow. An association between false positives and adverse responses to quality assurance questions (P = 0.04) illustrated the need for following NCCLS recommendations.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=262533Documentos Relacionados
- Comparison of Nucleic Acid Amplification Tests for Tuberculosis
- Evaluation of Three Nucleic Acid Amplification Methods for Direct Detection of Mycobacterium tuberculosis Complex in Respiratory Specimens
- Reliability of nucleic acid amplification for detection of Mycobacterium tuberculosis: an international collaborative quality control study among 30 laboratories.
- Performance Assessment of Two Commercial Amplification Assays for Direct Detection of Mycobacterium tuberculosis Complex from Respiratory and Extrapulmonary Specimens
- Comparison of Three Nucleic Acid Amplification Tests for Detection of Chlamydia trachomatis in Urine Specimens