Clinical laboratory comparison of a slide blood culture system with a conventional broth system.
AUTOR(ES)
Pfaller, M A
RESUMO
The recovery of bacteria and fungi from blood cultures was compared in conventional tryptic soy broth (TSB) bottles and in TSB bottles with an agarcoated slide attachment. A total of 2,662 sets of blood cultures, including 413 that were positive (15.5%), were evaluated. Significantly more gram-positive and gram-negative bacteria were recovered in the slide culture bottles than in conventional bottles (299 versus 253 isolates). Growth of gram-positive organisms and fungi was detected in the slide culture bottles 24 to 48 h earlier than in the TSB bottles. In addition, 76% of the isolates in the slide culture system were detected on the agar slide. In comparison, only 40% of the isolates in the TSB bottles were detected initially by blind subculturing. The incidences of contamination were 2.7% (71 cultures) for the slide culture system and 1.5% (39 cultures) for the TSB bottles.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=272402Documentos Relacionados
- Clinical comparison of an agar slide blood culture bottle with tryptic soy broth and a conventional blood culture bottle with supplemented peptone broth.
- Clinical evaluation of the lysis-centrifugation blood culture system for the detection of fungemia and comparison with a conventional biphasic broth blood culture system.
- Comparison of a slide blood culture system with a supplemented peptone broth culture method.
- Comparison of a radiometric and a broth-slide system for aerobic blood culture.