Comparison of Enzymatic and Microbiological Gentamicin Assays

AUTOR(ES)
RESUMO

The correlation coefficient between the rapid enzymatic and the overnight microbiological assays for 211 urine and serum specimens was 0.96. The 95% confidence limits yielded a correlation coefficient between 0.92 and 0.98. Both methods tended to underestimate the amount of a gentamicin added to urine. When only serum samples were considered, the predicted value obtained from the linear regression analysis of either method was within 0.57 μg/ml 99% of the time. This high degree of positive correlation will permit safe rapid adjustment of individualized patient gentamicin dosages.

Documentos Relacionados