Comparison of radioimmunoassay and fluorescent polarization immunoassay for quantitative determination of vancomycin concentrations in serum.
AUTOR(ES)
Ackerman, B H
RESUMO
A new fluorescent polarization immunoassay (Abbott Laboratories, Diagnostics Division, North Chicago, Ill.) was compared with a standard radioimmunoassay (American Diagnostics Corp., Newport Beach, Calif.) in 34 patients being treated with vancomycin. A total of 123 serum samples were divided and quantitatively analyzed for vancomycin by both assay methods. The results obtained indicated that the two assay methods are comparable (y = 1.01x - 0.81; r = 0.99).
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=270950Documentos Relacionados
- Comparison of enzyme-multiplied immunoassay technique with fluorescence polarization immunoassay for determination of gentamicin and tobramycin levels in serum.
- Comparison of a radioimmunoassay and a microbiological assay for measurement of serum vancomycin concentrations.
- Substrate-labeled fluorescent immunoassay for amikacin in human serum.
- Comparison of a semiautomated fluorescent immunoassay system and indirect immunofluorescence for detection of antinuclear antibodies in human serum.
- Comparison of fluorescence polarization immunoassay and bioassay of vancomycin.