Comparison of the Cathra Repliscan II, the AutoMicrobic system Gram-Negative General Susceptibility-Plus Card, and the Micro-Media System Fox Panel for dilution susceptibility testing of gram-negative bacilli.

AUTOR(ES)
RESUMO

A comparative evaluation was done to test the accuracy of the Cathra Repliscan II agar dilution system (Diagnostic Equipment, Inc., St. Paul, Minn.), the AutoMicrobic system with Gram-Negative General Susceptibility-Plus Card (Vitek Systems, Inc., Hazelwood, Mo.), and the Micro-Media Fox Panel micro broth dilution system (Micro-Media Systems, Inc., San Jose, Calif.) in determining MICs of 12 antibiotics for 200 gram-negative bacilli. Of the 200 strains tested, 12 isolates did not grow in one of the three systems. The 188 remaining organisms included 158 members of the family Enterobacteriaceae, 20 Pseudomonas spp., 5 Acinetobacter sp., 3 Aeromonas spp., and 2 Vibrio spp. A total of 2,256 organism-antibiotic combinations were analyzed for each system. An MIC was considered correct if two of the three systems were in agreement. When disagreements occurred, correct MICs were determined by the standard agar dilution method. With this criterion, overall agreements of the Cathra Repliscan II system, AutoMicrobic system, and Micro-Media Fox Panel system were 94.7, 94.9, and 95.5%, respectively. Tetracycline (20%), nitrofurantoin (20%), and ampicillin (16%) accounted for 56% of the discrepancies observed. These results indicate that all three systems perform with a high degree of accuracy for susceptibility testing of gram-negative bacilli.