Comparison of the Vitek Gram-Positive Susceptibility 106 Card, the MRSA-Screen Latex Agglutination Test, and mecA Analysis for Detecting Oxacillin Resistance in a Geographically Diverse Collection of Clinical Isolates of Coagulase-Negative Staphylococci
AUTOR(ES)
Yamazumi, T.
FONTE
American Society for Microbiology
RESUMO
The Vitek automated susceptibility testing system with a modified gram-positive susceptibility (GPS) 106 card (bioMerieux Vitek, Inc., Hazelwood. Mo.) and a rapid slide latex agglutination test (MRSA-Screen test; Denka Seiken Co., Ltd., Tokyo, Japan) were evaluated for their abilities to detect oxacillin resistance in coagulase-negative staphylococci (CoNS). The reference broth microdilution method and the detection of the mecA gene by PCR (“gold standard” reference result) were used to compare the results obtained with the commercial products. A total of 123 clinical isolates consisting of eight species were selected from U.S. surveillance collections. Among the mecA-positive isolates (95 strains), 30 isolates were initially negative on the MRSA-Screen test read at 3 min. When the agglutination reaction was extended for 10 min, 26 of the 30 isolates became positive. For a different four isolates, the oxacillin MIC was ≤0.25 μg/ml on the Vitek GPS 106 card. Among the mecA-negative isolates (28 strains), for two Staphylococcus warneri, two S. lugdunensis, and two S. saprophyticus strains MICs were ≥0.5 μg/ml by the reference broth microdilution method. Four of these isolates were also categorized as resistant with the Vitek GPS 106 card and two isolates were positive by the MRSA-Screen test. Overall, the MRSA-Screen test, GPS 106 card, and reference broth microdilution method had sensitivities of 95.7 (result at 10 min), 95.7, and 100%, respectively, and specificities of 92.8, 85.7, and 78.5%, respectively. Although the MRSA-Screen test required a slight procedural modification, both commercial methods achieved a sensitivity and specificity at detecting oxacillin resistance in CoNS at a level that was acceptable for clinical laboratory use.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=88400Documentos Relacionados
- Comparison of the Vitek Gram-Positive Susceptibility 106 Card and the MRSA-Screen Latex Agglutination Test for Determining Oxacillin Resistance in Clinical Bloodstream Isolates of Staphylococcus aureus
- Evaluation of a Latex Agglutination Test (MRSA-Screen) for Detection of Oxacillin Resistance in Coagulase-Negative Staphylococci
- Ability of the Modified Vitek Card To Detect Coagulase-Negative Staphylococci with mecA and Oxacillin-Resistant Phenotypes
- Evaluation of the Vitek Systems Gram-Positive Identification card for species identification of coagulase-negative staphylococci.
- Correlation of Oxacillin MIC with mecA Gene Carriage in Coagulase-Negative Staphylococci