Construção de uma cela de fluxo para medidas por espectrofotometria em fase sólida
AUTOR(ES)
Reis, Boaventura F., Rocha, Fábio R. P., Teixeira, Leonardo Sena Gomes, Costa, Antônio Celso Spinola, Korn, Mauro
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2000-02
RESUMO
A simple and low cost flow cell is proposed for measurements by solid-phase spectrophotometry employing a conventional spectrophotometer. The flow cell geometry allows the employment of a large amount of the solid support without causing both excessive attenuation of the radiation beam and increasing of the back-pressure. The adaptation of the flow cell in the optical path of the spectrophotometer in order to increase the precision is discussed. The flow cell characteristics were demonstrated by measurements of Co(II), employing 1-(2-tiazolylazo)-2-naphthol (TAN) immobilized on C18 bonded silica as solid support. The apparent molar absorptivity and coefficient of variation were estimated as 1.86 x 10(5) L mol-1 cm-1 and 1.4 % (n=15). A sample throughput of 40 determinations per hour and a detection limit of 15 mug L-1 (99.7 % confidence level) were achieved.