Controlled comparison of a new Becton Dickinson agar slant blood culture system with Roche Septi-Chek for the detection of bacteremia and fungemia.
AUTOR(ES)
Reimer, L G
RESUMO
Two commercially available agar slide blood culture systems, Septi-Chek (Roche Diagnostics, Div. Hoffmann-La Roche Inc., Nutley, N.J.) and VACUTAINER agar slant (Becton Dickinson [BD] Vacutainer Systems, Rutherford, N.J.), were compared in 5,628 paired blood cultures from adult patients. The systems were inoculated with equal volumes (5 ml) of blood. There was no statistically significant difference between the two systems in the overall recovery of clinically important microorganisms; however, significantly more Escherichia coli (P less than 0.05) organisms were detected in the Septi-Chek system. Concurrently, significantly more contaminants (P less than 0.001) also were present in Septi-Chek bottles. When microorganisms grew in both systems, there was no significant difference in the time to recovery of important organisms, except that anaerobes were detected more quickly in the BD system. This study suggests that the BD VACUTAINER agar slant is an acceptable alternative to the Septi-Chek system for routine blood cultures.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=267099Documentos Relacionados
- Comparison of Du Pont Isolator and Roche Septi-Chek for detection of fungemia.
- Clinical comparison of difco ESP, Wampole isolator, and Becton Dickinson Septi-Chek aerobic blood culturing systems.
- Controlled evaluation of BACTEC PLUS 27 and Roche Septi-Chek anaerobic blood culture bottles.
- Controlled evaluation of BACTEC Plus 26 and Roche Septi-Chek aerobic blood culture bottles.
- Comparative evaluation of the oxoid signal and Roche Septi-Chek blood culture systems.