Disc erosion in Models 103 and 104 of Beall mitral valve prostheses
AUTOR(ES)
Gómez, Ricardo
RESUMO
Three cases of severe disc variance and erosion of the Teflon-disc Beall mitral valve prosthesis (Models 103 and 104) are reported. In two patients, the Beall mitral valves were excised and replaced with two Björk-Shiley mitral valves. The remaining patient did not survive, and at autopsy, the lens was found at the aortic bifurcation level. Because of this potentially lethal complication, careful follow-up of patients with Beall mitral valve prostheses (Models 103 and 104) is recommended.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=287920Documentos Relacionados
- Ten-year pathological study of Beall mitral valve prostheses.
- Doppler echocardiography in the study of patients with mitral disc valve prostheses.
- Haemolysis related to mitral valve replacement with the Beall valve prosthesis
- Mitral valve replacement with ball valve prostheses