Distribution of bacteria and yeasts within the 10-ml Isolator during the processing of seeded blood samples.

AUTOR(ES)
RESUMO

Forty-five organisms consisting of stock cultures and clinical isolates of bacteria and yeast were separately inoculated into outdated blood bank blood to achieve a concentration of approximately 100 CFU/ml. Blood with each organism was introduced into groups of four Isolators (E. I. du Pont de Nemours & Co., Inc., Wilmington, Del.), which were then processed according to the Isostat instructions of the manufacturer. The supernatant, sediment, and wash (material removed from the surface of the slanted stopper after sediment removal) were inoculated onto 5% sheep blood agar plates. Cultures were incubated aerobically (5 to 10% CO2) at 35 degrees C for 48 to 72 h. From the 180 Isolators, the mean recovery was 6% (range, 0 to 48%) for the supernatant, 87% (range, 47 to 98%) for the sediment, and 8% (range, 3 to 23%) for the wash. Neither variation among technologists nor intentional misalignment of additional Isolators in the centrifuge could explain all of the losses of microorganisms from the sediment. The manual nature of the Isolator procedure, which led to the loss of significant amounts of organisms from the sediment, may help to explain false-negative Isolator results obtained from blood of patients, particularly when small numbers of pathogens are present.