Emprego de padrão único para a construção de curvas analíticas em sistema de análise por injeção em fluxo associado à espectrofotometria em fase sólida

AUTOR(ES)
FONTE

Química Nova

DATA DE PUBLICAÇÃO

2003-08

RESUMO

In this work two procedures were proposed for analytical curves construction using a single standard solution employing a flow injection system with solid phase spectrophotometric detection (FI-SPS). A flow cell contends the chromogenic reagent 1-(2-tiazolylazo)-2-naphtol was positioned on the optical path. The first procedure was based on controlled concentration of analyte on solid phase and the relations between absorbance and the total volume of injected allowed the calculation of analyte concentration. The second procedure was developed employing controlled dispersion/retention in flow system where analyte concentration was obtained by exploiting the relation between transient signals of samples and single standard solution at equivalent reading time. The procedures were successfully applied for zinc determination in synthetic solutions with good precision and accuracy at 95% confidence level.

Documentos Relacionados