Evaluation of Port-A-Cul transport system for protection of anaerobic bacteria.

AUTOR(ES)
RESUMO

The protection of anaerobes in Port-A-Cul (PAC) transport system (Bioquest, Div. of Becton, Dickinson &Co., Cockeysville, Md.) tubes and vials was studied. Ten species of obligately anaerobic bacteria commonly isolated from clinical specimens were used to prepare simulated swab and fluid specimens in high and low concentrations. Samples in PAC tubes and vials were held for 2, 24, and 48 h at ambient temperature and in a refrigerator. In addition, samples of the simulated specimens were exposed to controlled anaerobic and aerobic conditions in vented tubes and vials, with and without PAC medium, at ambient and refrigerator temperatures. Viable bacterial colony counts from specimens in PAC tubes and vials used as recommended by the manufacturer were consistently greater than those from specimens exposed to the different controlled conditions. The protection in PAC was about equal for specimens with either high or low concentrations of bacteria. Protection of the anaerobes in PAC was more obvious with swab than with fluid specimens. Quantitative recovery of anaerobes from refrigerated PAC samples, with few exceptions, was comparable to that from PAC samples held at ambient temperature.