Evaluation of the arterial blood pressure of dogs by two noninvasive methods.

AUTOR(ES)
RESUMO

The determination of the arterial blood pressure was done on 12 healthy mixed breed dogs in both the anesthetized and the conscious state, to evaluate two instruments (Doppler flow detector and infrasonde D4000), in their ability to indirectly determine arterial blood pressure. The coefficients of variation were higher with indirect methods when compared with the results obtained by cannulation. These coefficients were lower with the Doppler flow detector. The correlation study showed that both apparatuses were reliable in most situations. The infrasonde D4000 was more accurate than the Doppler in the conscious animals. However the results showed a lack of precision in hypertensive conscious dogs. The diastolic arterial blood pressure was particularly precise in the case of the anesthetized hypotensive dogs. Its sensitivity allowed it to register muscle movement artifacts. The Doppler flow detector showed less variation and was particularly accurate in both anesthetized and conscious hypertensive dogs. Its sensitivity allowed artifact movement sounds to be detected. The Doppler should be used in quiet surroundings or earphones should be worn by the evaluator. Some form of restraint is needed with the use of both instruments. Even if the correlations with the direct arterial blood pressure values were better with the infrasonde D4000, greater variations were found in the individual readings. The Doppler instrument represents in the hands of the investigators a better instrument for routine monitoring of blood pressure in the dog.