Evaluation of the autoSCAN-W/A rapid system for identification and susceptibility testing of gram-negative fermentative bacilli.

AUTOR(ES)
RESUMO

The autoSCAN-Walk-Away (W/A) system for identification and susceptibility testing was evaluated for 400 gram-negative fermentative bacteria by using the API 20E (366 isolates) and/or tube biochemical tests as the reference identification system and a frozen microdilution MIC tray system for susceptibility testing. The W/A system performed well for identification of this group of organisms representing 14 genera and 30 species, showing a sensitivity of 96% and results available in 2 h. Of the 16 misidentifications, 6 were with Serratia liquefaciens. A total of 63 isolates (17%) required further tests to complete the identification, compared with 106 (29%) of the isolates which required additional tests for the API 20E identification. Approximately half (32) of the additional tests with the W/A system were required in order to separate Citrobacter diversus from C. amalonaticus. For susceptibility determinations, the W/A system demonstrated an overall agreement of 93% (4,102 determinations) with 40 major errors (0.98%). However, of the 906 resistant organism-drug combinations in the study, there were 115 very major errors, for a false-susceptibility rate of 12.7% of the resistance determinations. Among these very major errors, 80% occurred with piperacillin and the cephalosporins. The W/A system completed the MIC determinations in 7 h; however, the difficulty in detecting resistance with some antimicrobial agents limited the advantages of the rapid susceptibility testing.

Documentos Relacionados