Evaluation of the pathotec Rapid I-D system for identification of Enterobacteriaceae.
AUTOR(ES)
Smith, P B
RESUMO
The PathoTec Rapid I-D System for identifying Enterobacteriaceae was evaluated with 471 cultures. In 4,910 individual test comparisons, 95.5% of the results agreed, with results of only two test strips, those for esculin hydrolysis and urease production, agreeing with conventional tests in less than 94% of the trials. The PathoTec system exhibited 94.3% accuracy in identifying these cultures in a double-blind study with conventional media and procedures as the alternate system. Two newly developed test strips, for 0-nitrophenyl-beta-D-galactopyranoside and ornithine decarboxylase, were found to be highly reliable.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=275092Documentos Relacionados
- Evaluation of the PathoTec „Rapid I-D System”
- Evaluation of the PathoTec „Rapid I-D System” and Two Additional Experimental Reagent-Impregnated Paper Strips
- Evaluation of the modified Micro-ID system for identification of Enterobacteriaceae.
- Evaluation of the MS-2 system for rapid identification of Enterobacteriaceae.