Evaluation of the Phenol Ammonium Sulfate Sedimentation Smear Microscopy Method for Diagnosis of Pulmonary Tuberculosis
AUTOR(ES)
Selvakumar, N.
FONTE
American Society for Microbiology
RESUMO
We compared the sensitivity and specificity of the phenol ammonium sulfate (PhAS) sediment smear microscopy method for detection of acid-fast bacilli with those of direct smear microscopy, using culture results for Mycobacterium tuberculosis as the “gold standard.” The sensitivities of the PhAS and direct smear methods were 85% (465 of 547) and 83% (454 of 547), respectively, and the specificity of each method was 97%. The PhAS method was better accepted by the laboratory technicians and safer but necessitates an overnight sedimentation, which delays reporting of results until 1 day after sputum collection.