Gamma radiation-sterilized, triple-lumen catheters coated with a low concentration of chlorhexidine were not efficacious at preventing catheter infections in intensive care unit patients.

In a randomized, double-blind trial, gamma radiation-sterilized, chlorhexidine-coated triple-lumen catheters were compared with uncoated control catheters for their ability to prevent catheter infection in 254 intensive care unit patients. The chlorhexidine coating was not efficacious, and a rabbit model demonstrated that reduction of chlorhexidine activity by gamma radiation sterilization was the likely explanation for the failure.