Incidence of Neisseria gonorrhoeae isolates negative by Syva direct fluorescent-antibody test but positive by Gen-Probe accuprobe test in a sexually transmitted disease clinic population.

AUTOR(ES)
RESUMO

To determine the accuracy of the Syva (Palo Alto, Calif.) direct fluorescent-antibody (DFA) test in comparison with the Gen-Probe (San Diego, Calif.) Accuprobe culture confirmation test, we tested 395 isolates of Neisseria gonorrhoeae from cultures obtained from patients attending a sexually transmitted disease clinic from 1 July 1991 through 30 June 1992. All isolates were tested for DFA reactivity with a polyclonal reagent (Difco Laboratories, Detroit, Mich.) and a monoclonal reagent (Syva, Inc., direct specimen test) and for specific molecular probe reactivity by the Gen-Probe Accuprobe culture confirmation test for N. gonorrhoeae. The 395 isolates gave positive results for the Gen-Probe culture confirmation test and the Difco polyclonal direct specimen test. However, 18 (4.6%) of the isolates were negative for N. gonorrhoeae by the Syva DFA test. With the exception of six beta-lactamase-positive isolates, all isolates that were negative by Syva DFA were sensitive to penicillin, tetracycline, spectinomycin, and ceftriaxone by disk-diffusion susceptibility testing. Auxotyping and serotyping studies indicated that strains negative by Syva DFA consisted of several variants. The frequency of N. gonorrhoeae isolates showing negative results by Syva DFA in this patient population ranged from 0 to 11.5%/month. Laboratories using only the Syva DFA test for confirmation of N. gonorrhoeae may incur a significant risk of misidentification.