Methicillin-resistant Staphylococcus aureus susceptibility testing with Abbott MS-2 system.
AUTOR(ES)
Carlson, J R
RESUMO
The antimicrobial susceptibilities of 100 methicillin=resistant Staphylococcus aureus strains were concurrently determined by the Abbott MS-2 System and by the standard disk diffusion method. Agreement between the two methods was 94% or greater for all of the antibiotics tested except for methicillin and gentamicin. This study indicates that the Abbott MS-2 cannot be relied upon for detection of methicillin resistance in clinical S. aureus isolates.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=181967Documentos Relacionados
- Reliability of the MS-2 system in detecting methicillin-resistant Staphylococcus aureus.
- Methicillin-Resistant Staphylococcus aureus Susceptibility Testing by an Automated System, Autobac I
- Susceptibility of methicillin-resistant Staphylococcus aureus to lysostaphin.
- Methicillin-resistant Staphylococcus aureus.