Overestimation of the Hepatitis C Virus RNA Content of Reference Preparations by the AMPLICOR HCV Monitor Test, Version 2.0
AUTOR(ES)
Pisani, Giulio
FONTE
American Society for Microbiology
RESUMO
An evaluation of the AMPLICOR hepatitis C virus (HCV) monitor test, version 2.0 (Roche Diagnostics), was carried out to investigate whether this test overestimates the HCV RNA content of reference preparations. Satisfactory accuracy was observed when the World Health Organization HCV international standard was included in the assay and a modified formula was used to calculate the viral content.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=154603Documentos Relacionados