Performance characteristics of bioassay, radioenzymatic assay, homogeneous enzyme immunoassay, and high-performance liquid chromatographic determination of serum gentamicin.

AUTOR(ES)
RESUMO

We compared the accuracy, precision, and between-method error of the microbiological assay, the radioenzymatic assay, the homogeneous enzyme immunoassay, and the high-performance liquid chromatographic assay for the quantitation of gentamicin in serum. Precision and accuracy were evaluated by reference samples prepared to contain 0.0 to 32.7 micrograms of gentamicin per ml. Correlations between the methods utilized patient sera with gentamicin concentrations ranging from 0.6 to 13.3 micrograms/ml. All methods were reliable within acceptable limits for routine clinical use; intermethod correlation coefficients exceeded 0.96. Relative to the microbiological assay, the alternative methods offer the advantage of rapid analysis. The elapsed times for acquiring data on a set of 10 specimens under routine operating conditions were 0.5 h by the enzyme immunoassay, 4 h by the radioenzymatic assay, 5 h by the high-performance liquid chromatographic assay, and 10 h by the microbiological assay.