Qualidade de medições em química analítica. Estudo de caso: determinação de cádmio por espectrofotometria de absorção atômica com chama
AUTOR(ES)
Chui, Queenie S. H., Zucchini, Ricardo R., Lichtig, Jaim
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2001-06
RESUMO
Calculation of uncertainty of results represents the new paradigm in the area of the quality of measurements in laboratories. The guidance on the Expression of Uncertainty in Measurement of the ISO / International Organization for Standardization assumes that the analyst is being asked to give a parameter that characterizes the range of the values that could reasonably be associated with the result of the measurement. In practice, the uncertainty of the analytical result may arise from many possible sources: sampling, sample preparation, matrix effects, equipments, standards and reference materials, among others. This paper suggests a procedure for calculation of uncertainties components of an analytical result due to sample preparation (uncertainty of weights and volumetric equipment) and instrument analytical signal (calibration uncertainty). A numerical example is carefully explained based on measurements obtained for cadmium determination by flame atomic absorption spectrophotometry. Results obtained for components of total uncertainty showed that the main contribution to the analytical result was the calibration procedure.
Documentos Relacionados
- DETERMINAÇÃO DE CÁDMIO E CHUMBO EM SANGUE POR ESPECTROMETRIA DE ABSORÇÃO ATÔMICA COM TUBO AQUECIDO NA CHAMA APÓS COMBUSTÃO INICIADA POR MICROONDAS
- COMBUSTÃO INICIADA POR MICROONDAS E DETERMINAÇÃO EM LINHA DE CÁDMIO E CHUMBO EM PLANTAS POR ESPECTROMETRIA DE ABSORÇÃO ATÔMICA COM TUBO AQUECIDO NA CHAMA
- Quantificação laboratorial de cobre sérico por espectrofotometria Vis comparável à espectrometria de absorção atômica com chama
- Pré-concentração de chumbo e cádmio em um sistema de micro extração líquido-líquido e determinação por espectrometria de absorção atômica com chama
- Determinação de ferro em sangue, soro e plasma por espectrometria de absorção atômica com chama.