Rapid determination of amphotericin B levels in serum by high-performance liquid chromatography without interference by bilirubin.
AUTOR(ES)
Hosotsubo, H
RESUMO
A high-performance liquid chromatographic method for the determination of amphotericin B concentrations in human serum without bilirubin interference was developed and compared with a microbiological assay. The high-performance liquid chromatographic assay utilized a reversed-phase trimethyl silica column, simple sample preparation, and visible detection. Reproducibility studies yielded coefficient-of-variation ranges from 1.02 to 2.11% for within-day precision and 2.88 to 4.32% for between-day precision. The correlation coefficient with the microbiological assay was 0.984 for amphotericin B.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=172355Documentos Relacionados
- Rapid determination of serum 5-fluorocytosine levels by high-performance liquid chromatography.
- Amikacin Assay in Serum by High-Performance Liquid Chromatography
- Performance characteristics of two bioassays and high-performance liquid chromatography for determination of flucytosine in serum.
- Identification of mycobacteria by high-performance liquid chromatography.