Reproducibility Problems with the Abbott Laboratories LCx Assay for Chlamydia trachomatis and Neisseria gonorrhoeae
AUTOR(ES)
Gronowski, Ann M.
FONTE
American Society for Microbiology
RESUMO
This study demonstrates that significant reproducibility problems can occur during routine use of the Abbott Laboratories LCx assay for Chlamydia trachomatis and Neisseria gonorrhoeae. These problems can go undetected by the quality control procedures outlined in the manufacturer's package insert. We outline here procedures for detecting and preventing contamination and reproducibility problems.