Sensitivity, precision, and accuracy of the Y1 adrenal cell enterotoxin assay.
AUTOR(ES)
Tenney, J H
RESUMO
The Y1 adrenal cell assay for heat-labile enterotoxin (LT) was found to be relatively insensitive for filtrates of toxigenic Escherichia coli H10407. When observations were made blindly and subjected to rigorous controls, reliable detection of LT occured only at filtrate dilutions from 1:4 to 1:10. Detection of LT was unreliable when E. coli H10407 was mixed with another enteric organism. Overall, 7.1% of observers' reports from a single assay were imprecise, and 2.6% would have resulted in errors in detection of LT; however, triplicate assays prevented false positive reports of LT detection. Routine testing for LT production in the clinical diagnostic laboratory awaits a simpler, more sensitive, and more direct method of testing stool specimens.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=272990Documentos Relacionados
- Accuracy, precision, and quality control of enzyme assays.
- Expression of the human apolipoprotein E gene suppresses steroidogenesis in mouse Y1 adrenal cells.
- The primacy of ultrasound in the assessment of muscle architecture: precision, accuracy, reliability of ultrasonography. Physiatrist, radiologist, general internist, and family practitioner's experiences
- Preparation, Sensitivity, and Specificity of Limulus Lysate for Endotoxin Assay
- Expression of murine 21-hydroxylase in mouse adrenal glands and in transfected Y1 adrenocortical tumor cells.