Simple inoculum standardizing system for antimicrobial disk susceptibility tests.
AUTOR(ES)
Barry, A L
RESUMO
A collaborative study was undertaken to evaluate a simple, convenient device which expedites inoculum standardization for antimicrobial disk susceptibility tests. The Inocupac system (Minnesota Mining & Manufacturing [3M] Co., St. Paul, Minn.) was used to perform disk tests in parallel with the standard Bauer-Kirby method. Five investigators tested 100 selected isolates, each in triplicate. Inter- and intralaboratory precision of both disk procedures was essentially comparable. The Inocupac system failed to consistently provide satisfactory growth with some streptococci, but when testing common gram-negative bacilli, staphylococci, and most enterococci, the Inocupac system gave zones about the same size (+/-2 mm) as the Bauer-Kirby procedure. Interpretive agreement between the two test systems varied from 90 to 99% with different antimicrobial agents, and repeated tests with the Bauer-Kirby method demonstrated the same degree of interpretive agreement. The Inocupac system is a valid alternative method for inoculating disk susceptibility tests.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=273292Documentos Relacionados
- Inoculum preparation for anaerobic susceptibility tests.
- Interpretive criteria and quality control limits for ceftibuten disk susceptibility tests. Collaborative Antimicrobial Susceptibility Testing Group.
- Evaluation of Mueller-Hinton agar for disk diffusion susceptibility tests.
- Levofloxacin disk potency and tentative interpretive criteria for susceptibility tests.
- Temafloxacin disk potency and tentative interpretive criteria for susceptibility tests.