Sistema multicomutado de análise em fluxo para determinação de amônio e monocloramina em águas residuais e produtos de desinfecção
AUTOR(ES)
Borges, Sivanildo da Silva, Acevedo, Maria Soledad Moura Soares Fernandez, Korn, Mauro
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2009
RESUMO
A binary sampling flow analysis system equipped with gas diffusion cell was developed for NH4+ and/or NH2Cl determination in wastewater and disinfection products samples based on the Berthelot reaction of the NH2Cl diffused through the semi-permeable PTFE membrane. The effect of the analytical conditions related to the reaction and flow parameters were evaluated and N-NH4+ and N-NH2Cl were determined in concentration ranges from 0.17 to 5 mg L-1 and from 0.5 to 14.5 mg L-1, respectively. Limits of detection (3σ) of 50 and 140 µg L-1 for N-NH4+ and N-NH2Cl were calculated, respectively, and RSD of 5 and 2% were calculated for 10 consecutive determinations of N-NH4+ (1 and 3 mg L-1) and N-NH2Cl (3 and 9 mg L-1), respectively with 30 determinations h-1.
Documentos Relacionados
- Estudos para a determinação eletroquímica de aril N-metil carbamatos totais em águas em sistema de análise por injeção em fluxo
- Determinação e especiação de carbono inorganico em aguas naturais e atmosfera por analise em fluxo
- Determinação de paracetamol em produtos farmacêuticos empregando um sistema de análise por injeção em fluxo com geração de ácido nitroso
- Determinação condutimetrica de nitrato, nitrito, amonio e ureia em analise por injeção em fluxo e construção de um analisador automatico
- PROCEDIMENTO AUTOMÁTICO PARA DETERMINAÇÃO ESPECTROFOTOMÉTRICA DE HIDROQUINONA EMPREGANDO MULTICOMUTAÇÃO EM SISTEMA DE ANÁLISE EM FLUXO