Studies of sensory conductions. Comparison of latencies of orthodromic and antidromic sensory potentials.
AUTOR(ES)
Murai, Y
RESUMO
Latencies of orthodromic and antidromic sensory potentials were compared in 16 normal median nerves. Peak latency was longer in the antidromic sensory response in every case. The peak latency of the antidromic sensory response depended on the distance between the recording electrodes; with shorter interelectrode distances it was similar to that of orthodromic sensory response.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=492185Documentos Relacionados
- Comparison of antidromic and orthodromic action potentials of identified motor axons in the cat's brain stem.
- Intracranial stimulation of the trigeminal nerve in man. III. Sensory potentials.
- The optimal recording electrode configuration for compound sensory action potentials.
- Orthodromic sensory action potentials from palmar stimulation in the diagnosis of carpal tunnel syndrome.
- Imaging mitochondrial membrane potentials.