Unexpected Errors in Gas Chromatographic Analysis of Methane Production by Thermophilic Bacteria
AUTOR(ES)
Kim, Byoung-Kwan
RESUMO
Unexpected errors in methane measurement by gas chromatography occurred when samples at thermophilic temperatures were analyzed. With a standard curve prepared at room temperature (25°C), stoppered bottles incubated and sampled at 37 to 85°C showed more methane upon analysis than bottles incubated at 25°C: values at 50, 63, and 85°C were 109, 126, and 125%, respectively, of the 25°C value. All variation between 4 and 50°C can be explained by the temperature difference between culture bottle and sampling syringe, and the variation of methane concentration can be predicted by the gas law. Between 50 and 63°C, there was a more dramatic rise than predicted by theory. These variations are important to consider if thermophilic methane production is to be measured accurately. Methods to avoid errors are discussed.