Use of the Treponema pallidum-specific captia syphilis IgG assay in conjunction with the rapid plasma reagin to test for syphilis.

AUTOR(ES)
RESUMO

The Captia Syphilis IgG enzyme immunoassay (EIA) was evaluated for use in conjunction with the rapid plasma reagin test (RPR) as a method to test for syphilis. A total of 1,288 serum specimens were tested by the routine laboratory protocol of the RPR followed by microhemagluttination assay for Treponema pallidum (MHA-TP) testing of RPR-reactive sera as well as the EIA-RPR protocol in which the automated EIA followed by a manual RPR test for EIA-positive specimens is used. When using the routine protocol, 131 specimens were initially reactive by the RPR, and 113 of these were reactive by MHA-TP. When using the EIA-RPR protocol, 170 specimens were initially positive by EIA, and of these, 112 were RPR reactive, indicating active disease. When compared to the routine protocol, the EIA-RPR protocol had sensitivity, specificity, and positive and negative predictive values of 96.5, 99.7, 97.3, and 99.7%, respectively. After resolution of discrepancies by additional testing, the adjusted sensitivity, specificity, and positive and negative predictive values were 100, 99.8, 98.3, and 100%, respectively. This evaluation demonstrates that when used in conjunction with the RPR, the Captia Syphilis EIA is a reliable method by which to test for syphilis.